During use on a patient, the ventilator alarms for errors 1111 and 1208, on two separate occasions.The unit gets restarted after each event, after the second restart the treatment continues without further incidents.The new input fails with the same errors.This time the unit is restarted after each event.After the second restart the treatment continues without further incidents.Event time sn/ (b)(4).1208,11:36.42pm,10-10-2022,o2unavailable.1111,11:36.42pm,10-10-2022,cbito2devicefailed.1208,02:15.32am,10-11-2022,o2unavailable.1111,02:15.32am,10-11-2022,cbito2devicefailed.
|