• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. 1.5T LINX, 13B; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. 1.5T LINX, 13B; ANTI-REFLUX IMPLANT Back to Search Results
Model Number LXMC13
Device Problems Device Appears to Trigger Rejection (1524); Detachment of Device or Device Component (2907); Patient Device Interaction Problem (4001)
Patient Problems Dysphagia/ Odynophagia (1815); Choking (2464)
Event Date 01/01/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Date sent: 11/9/2022.Only event year known: 2022.Exact implant date is unknown.Device was implanted in 2018.Assumed first month of year and first day of month.An analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.Lot number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what was the exact date of implant? when did the dysphagia symptoms begin? what was the date of the imaging which showed the discontinuous linx? if available, please share a copy of this imaging.Was the device initially effective in controlling reflux? were any events associated with the onset of symptoms (vomiting, retching, trauma, surgery)? did the patient undergo an mri since device implant? if so, when was the mri and what strength? did the patient have any other surgeries in the area? was any additional imaging performed since device implant? does the device appear to be in a continuous annular state in these images? we are interested in establishing a window when the device may have become discontinuous.Please share any additional images.What is the management plan? is device removal scheduled? is a replacement linx or fundoplication planned? when and if the explanation takes place can we ask that the procedure gets video recorded and the video shared? when and if the linx device is removed, may we ask that the device be returned for analysis? when using the linx sizing device what technique was used to determine the size? did the patient have an autoimmune disease? is the patient currently taking steroids / immunosuppressive drugs? did the patient have any pre-existing dysphagia or other conditions (other than gerd)? how severe was the dysphagia/odynophagia before intervention? were there any intra-operative complications during implant? was there any hiatal or crural repair done at the same time as the implant? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that the patients has experienced dysphagia and on imaging it shows that the device has broken.Patient experiencing difficulty swallowing and getting choked.Patient has had egd, manometry, dilation twice, barium swallows.Patient had a ct for a non-related issue which is when it was noticed that the device is broken and is in a u shape and the beads do not come together and it does not appear to have a "wire" attached between the beads.The patient has tried dilation which did not help so a ct and upper gi confirmed the broken device.
 
Manufacturer Narrative
(b)(4).Date sent: 12/19/2022.A manufacturing record evaluation was performed for the finished device batch number 15283, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 13B
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4188 lexington avenue north
shoreview MN
Manufacturer Contact
kate karberg
4188 lexington avenue north
shoreview 
3035526892
MDR Report Key15764238
MDR Text Key303395540
Report Number3008766073-2022-00226
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005332
UDI-Public00855106005332
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/12/2021
Device Model NumberLXMC13
Device Catalogue NumberLXMC13
Device Lot Number15283
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/28/2022
Initial Date FDA Received11/09/2022
Supplement Dates Manufacturer Received12/19/2022
Supplement Dates FDA Received12/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/12/2017
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-