• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG EASYPUMP; PUMP, INFUSION, ELASTOMERI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG EASYPUMP; PUMP, INFUSION, ELASTOMERI Back to Search Results
Model Number 4540018-02
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problems Nausea (1970); Hot Flashes/Flushes (2153)
Event Type  malfunction  
Event Description
As reported by the user facility: event 1 we have had some more issues with the easy pumps.Two more have finished early, like 24 hours.The pumps that are emptying too soon are the same ones we have had problems with cracking/leaking during filling them.Ii lt 270-54-s volume 270 ml, flow 5 ml/hr the medication in all cases is fluorouracil (chemotherapy) the patients got extremely nauseous, hot and flushed.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Manufacturer Narrative
Event 1: this report has been identified as b.Braun medical internal report number (b)(4).Two samples and one photo were received for further evaluation.The received samples and photo were taken to a visual inspection.Damages that would lead to a malfunction were not detected at one of the received samples.The sample with no defects was functionally tested and no damages were observed.Upon opening the top cap of the second sample, several cracks in the lli cone of the filling port were observed.The sample was leaked tested and failed.Incidents of this nature are due to failure in the production process.An approved project is in place to further address issues with leakage.A review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYPUMP
Type of Device
PUMP, INFUSION, ELASTOMERI
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM   34212
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15769552
MDR Text Key307223265
Report Number9610825-2022-00459
Device Sequence Number1
Product Code MEB
UDI-Device Identifier04046964448638
UDI-Public(01)04046964448638
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081905
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4540018-02
Device Catalogue Number4540018-02
Device Lot Number22B07GE561
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2022
Initial Date FDA Received11/10/2022
Supplement Dates Manufacturer Received10/25/2022
Supplement Dates FDA Received02/13/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-