AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2022
Event Type  malfunction  
Manufacturer Narrative
The subject device has been returned to olympus for evaluation.During troubleshooting, the customer was instructed to change the filter.It was noted, after the customer changed the filter, the reported issue was resolved.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus technical assistance center (tac), the cycle with the endoscope reprocessor was noted to be too long.It was indicated the amount of water supplied to the washing tank is weaker than before.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information (e1, e2, e3) received from the reporter.The event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the cycle issue could not be determined.It is possible that the water flow became weaker than before because the water filter was clogged with dirt.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿chapter 7 routine maintenance 7.2 replacing the water filter (maj-824) replace the water filter at least once a month to prevent contamination of the rinse water.Note ·filter life varies depending on a number of factors including incoming water quality and use volume.·using at least a prefilter (0.45 micron or less) can extend the life of the water filter.If the prefilter is properly installed, the water filter and the prefilter should be replaced at least once every 6 months.For information on the water pre-filtration system, contact olympus.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15772016
MDR Text Key307882082
Report Number9610595-2022-03778
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/12/2022
Initial Date FDA Received11/10/2022
Supplement Dates Manufacturer Received11/15/2022
12/22/2022
Supplement Dates FDA Received12/14/2022
01/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse