Catalog Number 400.834S
Device Problem Entrapment of Device (1212)
Patient Problem Foreign Body In Patient (2687)
Event Date 08/23/2022
Event Type  malfunction  
Event Description
Device report from synthes reports an event in china as follows: it was reported that this was a case from the health authority.During a procedure for excision of acoustic neuroma on (b)(6) 2022, when the surgeon inserted the screw, the screw was broken.In order to preserve the complete reduction of the bone flap, the screw head residue embedded in the bone flap could not be taken out.Another device was used to complete the surgery, and there were no adverse consequences to the patient.There was no surgical delay reported.This report involves one ti low profile neuro screw self-drilling 4mm.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device available for evaluation: complainant part is not expected to be returned for manufacturer review/investigation.The initial reporter is a j&j employee.Part number: 400.834s.Lot number: 674p138.Manufacturing site: (b)(4).Release to warehouse date: 25 february 2022.Expiry date: 01 february 2032.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI LOW PROFILE NEURO SCREW SELF-DRILLING 4MM
Type of Device
PLATE, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15779171
MDR Text Key307272111
Report Number8030965-2022-09648
Device Sequence Number1
Product Code JEY
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number400.834S
Device Lot Number674P138
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/19/2022
Initial Date FDA Received11/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age42 YR
Patient SexFemale
-
-