Model Number 5492VL |
Device Problems
Use of Device Problem (1670); Calibration Problem (2890); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/08/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the knobs on the pacing block of the patient cable would not turn.It was further reported that the cable was noted to be physically damaged and was not functioning correctly.There was no patient involvement.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis: analysis was able to confirm the customer comment that the knobs on the pacing block of the patient cables would not turn.Analysis was able to confirm the customer comment that the cables were noted to be physically damaged and were not functioning correctly.It was noted that the red knob assembly was loose from the lead connector and would not hold the cable due to damage.It was also noted that the knob was also damaged, tool marks were noted.It was further noted that the connector case back was damaged where the red knob assembly should be.Negative continuity tests were ok.Positive continuity tests were open due to the damage, as there was no knob to connect the cable to the lead connector.The following tests were completed with probes due to damage; continuity from the plug to positive on the lead connector was ok with no intermittent or shorted connections found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|