• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT FORTIFY ASSURA DR ICD, US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT FORTIFY ASSURA DR ICD, US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2359-40QC
Device Problems Failure to Capture (1081); High impedance (1291); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2022
Event Type  Injury  
Event Description
During follow-up, loss of sensing, loss of capture and increased pacing impedance were observed.The event was resolved by explanting and replacing the device.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of sensing, impedance, and capture anomaly was confirmed.Electrical testing revealed a hybrid anomaly.The hybrid anomaly was found to be the cause of the sensing, impedance, and capture anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15784800
MDR Text Key303548394
Report Number2017865-2022-45108
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Model NumberCD2359-40QC
Device Lot NumberP000074768
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/08/2022
Initial Date FDA Received11/11/2022
Supplement Dates Manufacturer Received02/24/2023
Supplement Dates FDA Received03/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-