Catalog Number 66801286
Device Problems Arcing (2583); Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, during a npwt, sparks were generated from the cord of a renasys tch class 2 power sply when the ac adapter was connected.The procedure was completed, without delay, using a s+n back-up device.Patient was not harmed.
 
Manufacturer Narrative
H3, h6: the device was returned for evaluation.A visual inspection was performed, the strain relief was found to be damaged as described.A documentation review was performed.The image supplied clearly shows that the strain relief has been damaged.The manufacturing records and processes for the reported batch contain no contributory factors, records confirm that this batch was manufactured, meeting the required specifications.All inspection and testing requirement was met the prior to release.Historical review details previous cases of this nature, which are considered isolated in scope and there are no open nor closed escalation actions.Related to this event type.The risk management files adequately captures both of the reported concerns, with no updates required.The probable cause is deemed to be that excess stress has been applied against the strain relief, by a user.No manufacturing, packaging, labelling, design, concerns nor adverse trend have been observed, therefore no corrective actions are deemed necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TCH CLASS 2 POWER SPLY
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15786517
MDR Text Key303841652
Report Number8043484-2022-00201
Device Sequence Number1
Product Code OMP
UDI-Device Identifier40565126630
UDI-Public40565126630
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66801286
Device Lot NumberC1713A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/19/2022
Initial Date FDA Received11/14/2022
Supplement Dates Manufacturer Received06/23/2023
Supplement Dates FDA Received06/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-