PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD10
Device Problems Failure to Power Up (1476); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2022
Event Type  malfunction  
Event Description
It has been reported to philips that the allura system lost power during a clinical procedure.No harm has been reported.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed the reported problem.The technical investigation was done and identified that there was issue with the mpdu (m-cabinet power distribution unit) control unit.The fse replaced the mpdu (m-cabinet power distribution unit) control unit, and the system was returned to use in good working order.The codes are updated based on the investigation outcomes.The device problem code was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15788280
MDR Text Key305091584
Report Number3003768277-2022-00942
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2022
Initial Date FDA Received11/14/2022
Supplement Dates Manufacturer Received10/19/2022
Supplement Dates FDA Received08/07/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-