• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX500H11C
Device Problem Degraded (1153)
Patient Problems Sore Throat (2396); Cancer (3262); Speech Disorder (4415); Nodule (4551); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/08/2022
Event Type  Injury  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient has alleged scraping of esophagus, trouble swallowing, voice change, polyps in adenoids and became malignant.The patient went to er, and spent 4 days in hospital.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer was previously contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient has alleged scraping of esophagus, trouble swallowing, voice change, polyps in adenoids and became malignant.The patient went to er, and spent 4 days in hospital.In previous report section b5 was incorrect, correct b5 should be - the patient alleged particles in the device.There was no patient harm or injury.
 
Manufacturer Narrative
The manufacturer previously submitted mdr 2518422-2022-94120-2 with incorrect sections b1, b2, h1, h6.Corrections to previous mdr are made in this report as follows.Section b1 was corrected to adverse event and product problem.(only product problem was checked in previous mdr).Section b2 was corrected to other serious or important medical events.(previously it was blank).Section h1 was changed from malfunction to serious injury.Section h6- health impact code was updated.
 
Manufacturer Narrative
The manufacturer previously reported an allegation of an issue related to sound abatement foam.Additional information was received and section b5 should be reported as: the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a continuous positive airway pressure (cpap) device's sound abatement foam.The patient alleged the patient has alleged scraping of esophagus, trouble swallowing, voice change, polyps in adenoids and became malignant.The patient went to er, and spent 4 days in hospital.There was no medical intervention required by the patient.The reported the patient has alleged scraping of esophagus, trouble swallowing, voice change, polyps in adenoids and became malignant.The patient went to er, and spent 4 days in hospital and its reported severity was reviewed by the manufacture's clinical expert.This event is assessed as not related to the device in this case.Based on the available information,the manufacture concludes no further action is necessary.There was no medical intervention required by the patient.The device has not yet returned to the manufacturer for evaluation.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.Section(s) b1, b2, has changed related to the complaint changing from the reported adverse event to a product problem.Section h1 has changed to reflect a malfunction.Section h6 health effect- impact code, type of investigation findings and investigation conclusions has been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key15791171
MDR Text Key303588274
Report Number2518422-2022-94120
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 06/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500H11C
Device Catalogue NumberDSX500H11C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/08/2022
Initial Date FDA Received11/14/2022
Supplement Dates Manufacturer Received11/18/2022
11/16/2022
11/16/2022
Supplement Dates FDA Received11/24/2022
04/09/2023
06/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/15/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-