The initial reporter stated there is a mismatch of patient data for one patient sample tested on the cobas u 411 analyzer.When manually running patient samples on the u 411, two test strips are tested sequentially on the analyzer without assigning a patient id to each test strip measurement.The first strip will not be read by the analyzer and then the results of the second strip are assigned to the patient/identifier of the first test strip.The customer provided data for one patient sample that was measured twice.The first measurement data was provided from the customer's host system as follows: color = yellow, density = 1015.000 (no units provided), ph = 9.0, nitrites = positive, leukocytes = 500.00 (no units provided), protein = 25.00 mg/dl, glucose = normal mg/dl, ketones = negative, urobilinogen = normal mg/dl, bilirubin = negative.The second measurement data was provided from a printout of the u 411 analyzer: density = 1.025 (no units provided), ph = 5, leukocytes = negative, nitrites = negative, protein = negative, glucose = normal, ketones = negative, uribilinogen = normal, bilirubin = negative, erythrocytes = negative.
|