Model Number CV-180
Device Problems No Display/Image (1183); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2022
Event Type  malfunction  
Manufacturer Narrative
In speaking with olympus technical assistance center (tac), the customer noted that they were using a third party sdi (serial digital interface) to hdmi (high-definition multimedia interface) converter and a samsung 55 inch consumer tv.The customer confirmed that both sdi outputs were outputting an image but could not confirm whether the converter was outputting a signal.At the time of the initial call the issue had not resolved, however upon a follow up call with tac the customer reported that the issue was resolved with no further details.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus technical assistance center (tac), the evis exera ii video system center was not connecting to the slave monitor.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Three attempts were performed to obtain additional information, but no response was received from the customer.The device was not returned; therefore, the reported phenomenon or condition of the device could not be confirmed.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The specific root cause of the reported problem could not be determined at this time because the device was not returned.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyamaodakura
nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15809775
MDR Text Key307575451
Report Number3002808148-2022-04249
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/18/2022
Initial Date FDA Received11/16/2022
Supplement Dates Manufacturer Received02/06/2023
Supplement Dates FDA Received02/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MAIN MONITOR; SAMSUNG 55 INCH TV.
-
-