LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Catalog Number 16-02-80 |
Device Problem
Corroded (1131)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/18/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in cambridge, united kingdom.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system3t was found to have a damaged cooling coil in the patient tank during cleaning.There was no patient involvement.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
On (b)(6) 2023, during follow up with the customer, it was clarified that the 3t heater cooler is currently being used as no malfunction has been experienced by the customer.The damaged coil is an aesthetic defect that does not have any impact on the functionality of the device.Based on this, the reportability evaluation of the event has been modified to not reportable since the issue has unlikely possibility to cause patient injury.
|
|