• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. OPT BLADELESS 12X100 STABILITY; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. OPT BLADELESS 12X100 STABILITY; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Model Number 2B12LT
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during an unknown surgery, it was difficult to get instruments in and out of the trocar.Another device was used to complete the surgery.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Date sent: 12/08/2022.D4: batch # x95d88.Investigation summary.The product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the 2b12lt device was received with the duckbill closed.Not allowing the obturator to be inserted trough the sleeve.No conclusion could be reached regarding what may have caused the reported incident.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch x95d88 number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPT BLADELESS 12X100 STABILITY
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15815547
MDR Text Key307592052
Report Number3005075853-2022-08028
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10705036014218
UDI-Public10705036014218
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122512
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2B12LT
Device Catalogue Number2B12LT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/01/2022
Initial Date FDA Received11/17/2022
Supplement Dates Manufacturer Received11/18/2022
Supplement Dates FDA Received12/08/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-