Model Number PM2134
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented for a pacemaker implant procedure.During the procedure, the physician could not feel the set screw on the pacemaker.The pacemaker was replaced during the procedure to resolve the issue.The patient was in stable condition throughout the procedure.
 
Manufacturer Narrative
The reported event of could not engage or tighten the atrial setscrew was confirmed.Final analysis found the pacemaker was returned.The atrial setscrew was present in its normal position in the connector block of the pacer.The user of the torque wrench during the implant procedure made numerous off-center and angled incorrect wrench insertions through the septum such that the setscrew was never being properly engaged.The problem was caused by the user¿s incorrect use of the torque wrench.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUSTAIN XL DC
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15819013
MDR Text Key303997569
Report Number2017865-2022-45750
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734504720
UDI-Public05414734504720
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model NumberPM2134
Device Catalogue NumberPM2134
Device Lot NumberP000133490
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2022
Initial Date FDA Received11/17/2022
Supplement Dates Manufacturer Received01/25/2023
Supplement Dates FDA Received01/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-