A falsely low d-dimer result was obtained on a patient sample on a sysmex cs-2500 system using innovance d-dimer reagent.The result was not reported to the physician(s) as the d-dimer result obtained from the patient five days prior had been elevated.The sample was repeated three times for d-dimer the following day using the same system and reagent, recovering higher.The higher results were reported, as the correct results, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely low d-dimer result.
|