Model Number AED PLUS |
Device Problem
Incorrect Interpretation of Signal (1543)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/25/2022 |
Event Type
malfunction
|
Event Description
|
Complainant alleged that while attempting to monitor a patient (age & gender unknown), the device issued a "shock advised" prompt for a heart rhythm they believe was non-shockable.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation for evaluation.The customer's report was not replicated or confirmed.The device was put through extensive testing without duplicating the report.An internal inspection of the device found no discrepancies.A review of the clinical file and all analyses from the event were "no shock advised" determinations.The device provides shock determination from ecg analysis only.The device is not advised to be used on a concious and alert patient as published in the administrator's guide.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|