AIZU OLYMPUS CO., LTD. BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-H1200
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This is not sold in the u.S.But a similar device is.The device is returned, and an evaluation completed for it.The user¿s complaint of leakage was confirmed.Upon inspection and testing, it was observed that the distal end rubber coating (a-rubber) was missing in places.The a-rubber adhesive was also detached.This is attributed to chemical or physical stress.Due to the missing a-rubber, there is leakage of the device.Other observations for the device are strange noise from the angle lever, and scratches found on insertion tube, operation part, switch box, rotating mechanism, up/down plate, and scope connector.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Event Description
Customer returned this device for evaluation and repair of a leakage issue.There is no reported harm to any patient or health care persons.Upon evaluation of the returned device, it was observed that the distal end rubber coating (a-rubber) was missing in places.The a-rubber adhesive was also detached.This is attributed to chemical or physical stress.This medwatch is being submitted for the reportable issue of the a-rubber being missing in places, as observed during device evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 year since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely due to external factors, or handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15823747
MDR Text Key303912200
Report Number9610595-2022-04204
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBF-H1200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/25/2022
Initial Date FDA Received11/18/2022
Supplement Dates Manufacturer Received12/07/2022
Supplement Dates FDA Received12/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-