This is not sold in the u.S.But a similar device is.The device is returned, and an evaluation completed for it.The user¿s complaint of leakage was confirmed.Upon inspection and testing, it was observed that the distal end rubber coating (a-rubber) was missing in places.The a-rubber adhesive was also detached.This is attributed to chemical or physical stress.Due to the missing a-rubber, there is leakage of the device.Other observations for the device are strange noise from the angle lever, and scratches found on insertion tube, operation part, switch box, rotating mechanism, up/down plate, and scope connector.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
|
Customer returned this device for evaluation and repair of a leakage issue.There is no reported harm to any patient or health care persons.Upon evaluation of the returned device, it was observed that the distal end rubber coating (a-rubber) was missing in places.The a-rubber adhesive was also detached.This is attributed to chemical or physical stress.This medwatch is being submitted for the reportable issue of the a-rubber being missing in places, as observed during device evaluation.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 year since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely due to external factors, or handling.Olympus will continue to monitor field performance for this device.
|