• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAEMONETICS PCS®2 PLASMA COLLECTION SYSTEM; PCS2,LIST NO,06002-110-EXP-EW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAEMONETICS PCS®2 PLASMA COLLECTION SYSTEM; PCS2,LIST NO,06002-110-EXP-EW Back to Search Results
Model Number 06002-110-EXP-EW
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Anemia (1706); Lethargy (2560)
Event Date 05/02/2022
Event Type  Injury  
Manufacturer Narrative
On (b)(6) 2022, a haemonetics field service engineer evaluated the device used by the donor.Diagnostics, functional test and visual inspections were performed and all systems passed.The device was found to have met manufacturer specifications with no malfunction found.The investigation shows there are no non-conformances against the device serial number used in the donation procedure and no corrective/preventive actions related to this complaint.A review of the device history record showed no issues during the manufacture of this device and all testing passed.The device was manufactured according to approved procedures and met all specifications for release, with no noted manufacturing deviations.A review of the service record for this device shows no similar complaints.There were no machine errors or issues with the disposables used during the donation procedure reported in the complaint.The pcs®2 plasma collection system collects blood product from a donor, then processes it through the system to separate plasma which is then expressed into a collection container.It is not possible to extract plasma directly from a donor as stated in the complaint.There is no evidence to suggest that the donor reaction was related to the device used during the plasmapheresis procedure.
 
Event Description
Letter received from donor's representative on (b)(6) 2022 alleging: (i) donor was injured while attending a donation procedure at the center, (ii) the device failed to alert the technician that it was primarily drawing plasma from the donor.The technician noticed the discrepancy in the tubing coloring, which was allegedly filled with transparent plasma from donor and not blood and plasma.Allegedly this procedure went on for quite some time to the point where uncharacteristic lethargy was present and donor wanted to do nothing but sleep.Donor was removed from device and given energy replenishing libations.Donor was advised not to return to donate for 90 days for the blood cells to revive.After this occurrence, donor was rushed to the hospital for chronic lethargy and was diagnosed as being severely anemic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PCS®2 PLASMA COLLECTION SYSTEM
Type of Device
PCS2,LIST NO,06002-110-EXP-EW
Manufacturer (Section D)
HAEMONETICS
125 summer street
boston MA 02110
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer Contact
brenda bruyere
125 summer street
boston, MA 02110
MDR Report Key15824522
MDR Text Key303902795
Report Number1219343-2022-00049
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK060033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number06002-110-EXP-EW
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/20/2022
Initial Date FDA Received11/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/04/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other; Hospitalization;
Patient SexMale
-
-