Catalog Number 8065753109
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported transparent tube connecting the cutter came off during a vitrectomy surgery.The surgery was completed after replacing the product with another one.There was no patient harm.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Upon visual inspection it was confirmed that white-lined tube was detached from the cutter probe port of the engine.The root cause is consistent with a manufacturing error during probe assembly where adhesive was not properly applied on the manifold.After an investigation of this complaint, it has determined that no further actions will be pursued at this time.Quality in-process controls are in place to assist in mitigating this issue from occurring.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15832576
MDR Text Key307666501
Report Number1644019-2022-00964
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657531097
UDI-Public00380657531097
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number8065753109
Device Lot Number2539745H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/26/2022
Initial Date FDA Received11/21/2022
Supplement Dates Manufacturer Received01/25/2023
Supplement Dates FDA Received02/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
-
-