MAUDE Adverse Event Report: ZIMMER BIOMET, INC. BONE SCREW SELF-TAPPING 6.5 MM DIA. 20 MM LENGTH; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. BONE SCREW SELF-TAPPING 6.5 MM DIA. 20 MM LENGTH; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2022
Event Type  malfunction  
Event Description
It was reported that after a hip procedure, the surgeon noted on an x-ray that the screw had gone through the screw hole of the cup.At this time, the patient is not experiencing any harm and the surgeon does not plan to do any replacements.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
(b)(4).Concomitant medical products: 110010265 ¿ g7 cup ¿ 7195151.Report source japan.The product will not be returned to zimmer biomet for the investigation as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: anatomic alignment of the right hip arthroplasty.One of the acetabular fixation screws extends entirely through the screw hole into the acetabulum.Osteopenia.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW SELF-TAPPING 6.5 MM DIA. 20 MM LENGTH
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15833595
MDR Text Key307665440
Report Number0001822565-2022-03275
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K934765
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00625006520
Device Lot NumberJ7288433
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/27/2022
Initial Date FDA Received11/21/2022
Supplement Dates Manufacturer Received12/19/2022
Supplement Dates FDA Received12/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-