MAUDE Adverse Event Report: BOSTON SCIENTIFIC DE COSTA RICA S.R.L. POLARSHEATH; VASCULAR GUIDE CATHETER, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC DE COSTA RICA S.R.L. POLARSHEATH; VASCULAR GUIDE CATHETER, SINGLE-USE Back to Search Results
Device Problems Use of Device Problem (1670); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2022
Event Type  malfunction  
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event Description
It was reported that during a cryo-ablation procedure using a polarsheath air was see in the sheath's side port after aspirating with the balloon catheter inside of the sheath.They replaced the sheath and repeated the aspiration with the balloon catheter inside the second sheath, air was once again observed.They then removed the catheter and re-inserted the catheter without applying the negative pressure and no air was observed.The procedure was completed with the original catheter and the second sheath with no patient complications.At no point was any air seen in the patient.The sheath is not expected to be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARSHEATH
Type of Device
VASCULAR GUIDE CATHETER, SINGLE-USE
Manufacturer (Section D)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway
global park, la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key15835160
MDR Text Key307561471
Report Number2124215-2022-47713
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2022
Initial Date FDA Received11/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
POLARSHEATH
-
-