It was reported that during a cryo ablation procedure, the ablations were ineffective.Due to the patient's anatomy, it was difficult to maneuver the catheter without perforating the artery. the catheter length also contributed to the decision to switch the case to radiofrequency and the physician wanted to perform left side bypass ablation. the case was completed with radiofrequency. no patient complications have been reported as a result of this event.
|
Product event summary: the 217f3 electrophysiology (ep) catheter with lot 13254 was returned and analyzed.During external visual inspection of the shaft segment, a kink was observed on the shaft.The catheter smart chip data was reviewed.Data indicated that the catheter was used for 24 applications on the reported event date.During system performance testing with the test console, the console terminated the application and generated system notice 50005 (indicating that the safety system detected fluid in the catheter and stopped the injection).During pressure testing and inspection of the shaft segment, a shaft breach was observed at shaft.In conclusion, the ep catheter failed the returned product inspection due to a shaft kink and breach.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|