Due to system limitations, only one initial reporter name was addressed in as (b)(6).As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer.The investigation of the reported event is currently underway.(expiry date: 11/2022).
|
It was reported that during an angioplasty procedure, the pta balloon allegedly failed to deflate.It was further reported that the balloon was punctured with the tail end of the extra stiff wire.Reportedly, the balloon was allegedly difficult to remove.There was no reported patient injury.
|
H10: due to system limitations, only one initial reporter name was addressed in e1 as paola quintilii.Other two initial reporters are thomas pilgrim and martin glöckler.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one atlas gold pta dilatation catheter was received for evaluation.During visual evaluation, no kinks were noted to the catheter shaft and no anomalies to luers/y-body.During functional testing, the balloon was cut.A kink was noted to inner guide-wire lumen and glue bullet was not seated in the correct position.No other functional testing was performed.Therefore, the investigation is inconclusive for both the reported deflation problem and balloon removal difficulty as functional testing could not be performed due to the condition of the device.A definitive root cause for the reported deflation problem and balloon removal difficulty could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 11/2022),.
|