A falsely low free light chains, type lambda (flc lambda) result was obtained on a patient sample run in a 1:20 dilution on bn ii system using n latex flc lambda reagent.The sample was then repeated for flc lambda using a 1:5 manual sample dilution, also recovering low.The sample was then repeated for flc lambda using a 1:400 manual sample dilution, recovering higher.The sample was then repeated for flc lambda using a 1:2000 manual sample dilution, recovering higher and fitting the patient's electrophoresis result.The flc lambda result obtained with the 1:2000 dilution was reported, as the correct result, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely low flc lambda results.
|