Catalog Number SLC2012X
Device Problem Burst Container or Vessel (1074)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
An attempt was made to use one solarice rx ptca balloon catheter to treat a mildly tortuous, moderately calcified lesion in the mid right coronary artery (rca) with 95% stenosis.It was reported that the balloon burst when the device was in the patient.The patient is alive with no injury.Please note that this device (solarice rx) is not marketed in the united states; however, the device is deemed the same as the united states marketed device (euphora rx).
 
Manufacturer Narrative
Image analysis: video showing the pressurisation of the balloon of the device.Liquid is confirmed to be exiting the distal end of the balloon, with a spray of liquid exiting through the leak point.The balloon appears to be leaking through a point on the balloon, rather than a balloon bust having occurred.Additional information: the device was inspected with no issues.Negative prep was performed with no issues.Resistance was not noted while advancing the device to the lesion.The device was being used to pre-dilate the lesion.The balloon burst occurred after 2 inflations at 12 atm.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis: the device was received for analysis.A kink was evident to the hypotube.The device returned with balloon folds expanded.Deformation was evident to the distal tip.The balloon failed negative prep.On pressurisation of the device, liquid was observed exiting the balloon working length, indicating a pin hole tear was present on the balloon material, on the balloon working length.The balloon failed to maintain pressure.No other damage evident to the remainder of the device.Additional information: annex b, annex c <(>&<)> annex d codes medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLARICE RX
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY,
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15850428
MDR Text Key306930399
Report Number9612164-2022-04368
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K143480
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSLC2012X
Device Lot Number224550589
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/10/2022
Initial Date FDA Received11/23/2022
Supplement Dates Manufacturer Received11/23/2022
01/17/2023
Supplement Dates FDA Received12/14/2022
01/25/2023
Date Device Manufactured06/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
-
-