• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. GYNECARE TVT OBTURATOR; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. GYNECARE TVT OBTURATOR; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Model Number 810081
Device Problem Unraveled Material (1664)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/28/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested and the following was obtained: what is the status of the mesh? mesh is still implanted in the patient.It was after the procedure was completed that the surgeon examined the left over mesh to identify that it was fraying, falling apart at the edges.Small pieces were falling off as she stretched it out across a raytec.Patient current status/outcome? response from medical advisor: i have spoken with surgeon and the situation is currently resolved.The patient is not showing any signs of an adverse outcome.She will not be removing the product at this stage and will be monitoring the patient every 6 months for surveillance.Just checking if there any mesh parts are available for return? there is no mesh parts available for return.They were discarded.To date it has been reported that the device will not be returned.If the device or further details are received at a later date a supplemental medwatch will be sent.(b)(4).
 
Event Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and mesh was implanted.After the procedure was completed, the surgeon examined the left over mesh and identified that it was fraying and falling apart at the edges.Small pieces were falling off as she stretched it out across a raytec.The patient is not showing any signs of an adverse outcome and the product will not be removed.Additional information was requested.
 
Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.
 
Manufacturer Narrative
(b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6 photo analysis summary: neuchâtel team received for evaluation one photo of tvto device (product code 810081), batch 3941399.It can be observed in the photo that the device was manipulated.It can be also observed in the photo that the mesh is frayed with several particles around, and damaged.Additionally, organic matter can be seen around.The defect seen during the photo evaluation is aligned with the defect described in the event description, however, the defect identified is not linked to a manufacturing issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT OBTURATOR
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits godet 20
neuchatel 2000
SZ   2000
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15852351
MDR Text Key307794133
Report Number2210968-2022-09756
Device Sequence Number1
Product Code OTN
UDI-Device Identifier10705031000346
UDI-Public10705031000346
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K033568
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model Number810081
Device Catalogue Number810081
Device Lot Number3941399
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/02/2022
Initial Date FDA Received11/23/2022
Supplement Dates Manufacturer Received01/17/2023
01/16/2023
Supplement Dates FDA Received01/17/2023
01/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexFemale
-
-