A physician was using amphirion deep for the treatment of an 80mm fibrous lesion with chronic total occlusion in the posterior tibial artery.The device was prepped as per the ifu with no issues identified.A non-mdt inflation device was used and nacl plus contrast medium as inflation fluid.Two balloons were used in the procedure.It was reported that the first balloon had difficulties with inflation, first to nominal after to burst and after 20 bar still no contrast to see.Pressure went back to 12 bar and deflate the balloon was not possible, so they had to puncture through the skin to get the balloon out the patient.Afterwards they used a second balloon same size.No problems with inflation for this balloon but with deflation direct blood was in the catheter and inflation device but the contrast stayed in the balloon.After a while some contrast slowly came out and they could get the balloon out of the patient.No further patient injury reported for this event.
|
Product analysis an attempt to inflate the balloon was unsuccessful and a microscopic inspection found that the balloon bond appeared to be stretched.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|