The microsensor (id (b)(4)) was returned for evaluation.Device history record (dhr) - lot 3943845 (sn (b)(6)) met specifications when released.Failure analysis - the issue of the complaint was not confirmed.No visible damage to the millar sensor, catheter material, or connector.The icp express read 492.The device passed electronic, noise, linearity/hysteresis, and signal drift tests.The root cause of the issue reported by customer could not be determined as the device worked correctly.However, the possible root cause of the defect reported by the customer could be due to excessive pressure applied to product.
|