Model Number KNEE SCORPION |
Device Problem
Break (1069)
|
Patient Problems
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
|
Event Description
|
On (b)(6) 2022, it was reported by a sales representative via email that an ar-12990 knee scorpion top jaw broke off inside the patient's joint.This was discovered during a meniscal root repair procedure.Surgeon was able to remove all broken fragments.
|
|
Manufacturer Narrative
|
Complaint confirmed.One unpackaged ar-12990 knee scorpion serial/batch number (b)(6) was received for investigation.Functional testing found that the device does not function as required.Visual evaluation found that the device was broken off at the distal end missing the top jaw.The fragments generated during the event were not returned for analysis.The observed condition is most likely the result of applying excessive force while grasping and manipulating the suture.
|
|