• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ALLIANCE II; SYRINGE, BALLOON INFLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ALLIANCE II; SYRINGE, BALLOON INFLATION Back to Search Results
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2022
Event Type  malfunction  
Manufacturer Narrative
Approximated based on the date the manufacturer became aware of the event.The complainant was unable to provide the upn and lot number.Therefore, the manufacture and expiration dates are unknown.(b)(4).
 
Event Description
Note: this report pertains to a cre fixed wire dilatation balloon and an alliance inflation syringe used during the same procedure.It was reported to boston scientific corporation that a cre fixed wire dilatation balloon and an alliance inflation syringe were used in the esophagus during an esophagogastroduodenoscopy (egd) with dilation procedure on an unknown date.During the procedure, the pressure gauge kept dropping to 1 or 2 when the handle was released.The user continued inflation and over inflated the balloon.The balloon popped but did not break into multiple pieces and was removed from the patient.The customer stated that no pieces of the balloon detached inside the patient.The procedure was completed at this time.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLIANCE II
Type of Device
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15877403
MDR Text Key307707449
Report Number3005099803-2022-06960
Device Sequence Number1
Product Code MAV
UDI-Device Identifier08714729129332
UDI-Public08714729129332
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/03/2022
Initial Date FDA Received11/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-