Model Number 9735602 |
Device Problem
Human-Device Interface Problem (2949)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/11/2022 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding a navigation system being used intraoperatively during a functional endoscopic sinus surgery (fess).It was reported that the system registered the patient wrong.There was a surgical delay of less than one hour.There was no impact on the patient outcome.Additional information was received.It was reported that registration was not accurate as it was the wrong computed tomography (ct), therefore different anatomy.
|
|
Manufacturer Narrative
|
Concomitant medical products: section information references the main component of the system.Other relevant device(s) are: product id: 9735638, version #: 1.0.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H2/h6: the system was serviced, no components were replaced and the system passed all testing.Codes b01, c19, and d14 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3, h6: a software analysis was initiated.However, the software evaluation found that a probable cause was unable to be determined.Codes b01, c19, d15 are applicable to this analysis medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|