PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10/10; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD10/10
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/21/2022
Event Type  malfunction  
Event Description
It has been reported to philips that, system was unable to power on.No harm has been reported to philips.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.A philips field service engineer (fse) inspected the system onsite and confirmed that the system shuts off.Fse checked the input power supply, front pdu, crcb all found ok.Fse then checked the fuses on mpdu and found 500ma fuse busted.Fse replaced the damaged fuse in the mpdu.After replacement, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10/10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15880619
MDR Text Key305270801
Report Number3003768277-2022-01252
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeRP
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10/10
Device Catalogue Number722005
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/21/2022
Initial Date FDA Received11/29/2022
Supplement Dates Manufacturer Received12/06/2023
Supplement Dates FDA Received12/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/18/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-