• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD10
Device Problems Appropriate Term/Code Not Available (3191); Radiation Output Failure (4027)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2022
Event Type  malfunction  
Event Description
Issue: malfunction / login / patient on table (planned diagnostics).Description of the problem: filming not possible.Fm: not enough storage space, delete study.It has been reported to philips that the filming of the image was not possible and the procedure was deleted.No harm has been reported.The available information does not indicate whether the device was in clinical use.To date, no further information was received.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was in clinical use for a diagnostic procedure, which could be completed as planned.Additionally, it was clarified that initial problem statement did not reflect the issue experienced by the customer.The customer thought that the entrance dose limitation (edl) in the system's database was higher than 10r.A philips field service engineer inspected the system on site and confirmed that this was not the case and that the system was operating according to specification.After performing some configuration changes requested by the customer, the system was returned to use in good working order.The codes were updated based on the investigation outcome.Device problem code was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15881539
MDR Text Key305274574
Report Number3003768277-2022-01255
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/23/2022
Initial Date FDA Received11/29/2022
Supplement Dates Manufacturer Received03/24/2023
Supplement Dates FDA Received09/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-