• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR Back to Search Results
Model Number 00-3020
Device Problems Therapeutic or Diagnostic Output Failure (3023); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2022
Event Type  malfunction  
Event Description
During preventive maintenance, in tempus ls, a dpm fault was populated when the device was being tested for pacer outputs.The device didn't give output in the rage of the set pacer.The error created a banner that said ¿dpm fault.¿ a user report was received related to a reported product problem which is currently being investigated.At the time of reporting, the device has not yet been returned for investigation.Further updates will be provided when the device is received, and the investigation is in progress.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS-MANUAL
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough 
UK  
MDR Report Key15886843
MDR Text Key304894337
Report Number3003832357-2022-00059
Device Sequence Number1
Product Code LDD
UDI-Device Identifier07613365002737
UDI-Public7613365002737
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/01/2022
Initial Date FDA Received11/30/2022
Supplement Dates Manufacturer Received02/21/2023
Supplement Dates FDA Received05/22/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-