Catalog Number EMC9601N |
Device Problems
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/13/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that nine (9) buretrol blood administration sets were leaking from the heatseal chambers due to cuts/holes in the heatseal welds.This was observed during leak testing.It was noted that three (3) sets leaked at 50 kpa for the 15 seconds testing.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|