Model Number LEG ATTACHMENT PAD FOR COROMETRICS FETAL SCALP ELECTRODE CABLE
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/10/2022
Event Type  Injury  
Manufacturer Narrative
At this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.On follow-up, the customer confirmed that they attached/held down the leg plates and connection points via medical tape.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
It was reported to vyaire medical that the 2464aao-20 - leg attachment pad for corometrics fetal scalp electrode cable are not sticking to the legs and are having to be taped down.Also, the actual leads are not clipping onto the fse (fetal scalp electrode) attachments.The event occurred during patient use.The customer confirmed that there was no patient harm associated with the reported event.
 
Manufacturer Narrative
Result of investigation: the device history paperwork was reviewed for all lots, specifically reviewing the foam and snap lots to determine what lots were used and when.There were multiple lots of both the snaps and foam used during production.Actual suspect sample was not returned, but investigation was done using a different lot number with the same complaint returned from the user.Visual inspection reveals some of the snaps were broken.The wire connector was mated with the returned samples.The connector/snap mating found that some of the snaps were difficult to press into the connector.However, no electrodes disconnected from the device.Also, peel testing could not be performed on the foam of the returned samples as the strip label was impacting the adhesion results, skewing them to a much higher result.Therefore, the reported failure could not be duplicated/reproduced, and the root cause could not be determined.Though, the broken snap may contribute to the connector and electrode mating issue.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITAL SIGNS¿ LEG ATTACHMENT PAD
Type of Device
ELECTRODE, ELECTROCARDIOGRAPH
Manufacturer (Section D)
VYAIRE MEDICAL OY
kuortaneenkatu 2 helsinki ete
la-suomen laani, finland fin
helsinki
FI 
Manufacturer (Section G)
GRAPHIC CONTROLS ACQUISITION CORP.
400 exchange st
buffalo NY
Manufacturer Contact
erika bonilla
510 technology drive
irvine, CA 92618
MDR Report Key15888235
MDR Text Key304552700
Report Number3010838917-2022-00030
Device Sequence Number1
Product Code DRX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLEG ATTACHMENT PAD FOR COROMETRICS FETAL SCALP ELECTRODE CABLE
Device Catalogue Number2464AAO-20
Device Lot Number313594
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/31/2022
Initial Date FDA Received11/30/2022
Supplement Dates Manufacturer Received03/21/2023
Supplement Dates FDA Received04/04/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-