• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION BIPAP AUTOSV; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION BIPAP AUTOSV; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number DSX900H11C
Device Problem Degraded (1153)
Patient Problem Insufficient Information (4580)
Event Date 11/18/2022
Event Type  malfunction  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.There was no report of serious patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION BIPAP AUTOSV
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key15892627
MDR Text Key305540336
Report Number2518422-2022-100721
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX900H11C
Device Catalogue NumberDSX900H11C
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/18/2022
Initial Date FDA Received12/01/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/16/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberRES 88058
Patient Sequence Number1
-
-