STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HARH36
Device Problem Fracture (1260)
Patient Problem Insufficient Information (4580)
Event Date 10/01/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, evidence of clinical use was identified.The blade was fractured and separated from the device tip.Inspection of the fractured blade revealed signs of gouging and outer coating erosion as a result of activating against solid/hard materials.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause(s) are: too much force or torque applied to instrument, or grasping/pulling; incidental and prolonged activation against solid surfaces, such as bone or plastic; and/or attempting to bend, sharpen, or otherwise alter the shape of the blade.The instructions for use (ifu) state: use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice, indicating that sufficient torque has been applied to secure the blade; do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury; avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.Note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.Note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post market surveillance.
 
Event Description
It was reported that a stryker device was sent to cg labs (ethicon).The complainant is not aware of any patient injury, medical intervention, or extended procedure time.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key15894087
MDR Text Key306248409
Report Number0001056128-2022-00042
Device Sequence Number1
Product Code NLQ
UDI-Device Identifier07613327346251
UDI-Public07613327346251
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K161693
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHARH36
Device Catalogue NumberHARH36RR
Device Lot Number14014173
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/11/2022
Initial Date FDA Received12/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-