• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SINGAPORE DTX SAFEDRAW KITS; COMPUTER, BLOOD-PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SINGAPORE DTX SAFEDRAW KITS; COMPUTER, BLOOD-PRESSURE Back to Search Results
Catalog Number 689549/JPA
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2022
Event Type  malfunction  
Event Description
The account alleges that during a hemodynamic pressure monitoring procedure, the clinical staff noted that the pressure monitoring set was leaking fluid from the 3-way stopcock.The device was replaced for the patient.No additional patient consequences to report.
 
Manufacturer Narrative
The suspect device is expected to return for investigation.A follow-up report will be submitted once the investigation is complete.
 
Manufacturer Narrative
The suspect device was returned for evaluation.The complaint was confirmed and the root cause is attributed to clinical use and overtightening during use.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DTX SAFEDRAW KITS
Type of Device
COMPUTER, BLOOD-PRESSURE
Manufacturer (Section D)
MERIT MEDICAL SINGAPORE
198 yishun ave 7
singapore, 76892 6
SN  768926
Manufacturer (Section G)
MERIT MEDICAL SINGAPORE
198 yishun ave 7
singapore, 76892 6
SN   768926
Manufacturer Contact
bryson heaton bsn,rn.
1600 merit parkway
south jordan, UT 84095
MDR Report Key15894761
MDR Text Key305836965
Report Number8020616-2022-00113
Device Sequence Number1
Product Code DSK
UDI-Device Identifier00884450407370
UDI-Public884450407370
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number689549/JPA
Device Lot NumberC1819196
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/22/2022
Initial Date FDA Received12/01/2022
Supplement Dates Manufacturer Received12/16/2022
Supplement Dates FDA Received12/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-