This event has been reported by the importer on mdr# 2429304-2022-00146.The device referenced in this report was returned to olympus for evaluation.Preliminary findings are reported.Physical evaluation of the returned device: visual inspection as received condition under a microscope found multiple scratches, scrape marks on the distal end.There was a bubble present in propagation fluid medium, but no medium leakage, and the flexible shaft appeared to be slightly stretched.Also, found a small pin hole and multiple scrape marks along the insertion tube.The damaged section of the working length was measured approximately 580mm from distal end.Observed that there are no abnormalities on the connector section.The ultrasound probe was then connected properly into the olympus test probe driving unit maj-1720 and checked with an endoscopic ultrasound center me-2.The frequency detection was working correctly and displayed 20mhz on the monitor screen.During initial functional inspection, the us image did not appear on the screen due to bubble present; however, after vigorously shaking the distal end until all air bubble had been eliminated from inside the transparent sheath section of the ultrasonic probe, the us image displayed on screen, but appeared to be blurry, and low resolution.The damage found on the working length is likely due mishandling.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
|
This supplemental report is being submitted to correct the previously reported initial mdr, based on additional information, based on legal manufacturer's investigation.It was determined that the patient's outcome is unrelated to the use of the device, and that the device reportedly involved into this reported event did not malfunction.
|