SMITH & NEPHEW, INC. PCH BSKT DCKBL UPBTR 1.5MM; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 012014
Device Problems Flaked (1246); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
It was reported that during a knee scope, the punch failed to performed as expected.It is unknown if the device would not bite through tissue or it was difficult to open and close and was shedding metal.The procedure was completed with a s+n back up device.There was a non-significant surgical delay and no further complications were reported.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection of the returned device found that it is not in its original packaging.The device has discoloration throughout the metal and there is some gouging of the metal on the shaft but there is no material missing.The cutting surface shows wear and damage from use but there is no material missing.A functional evaluation of the device found that it was able to cut cleanly through the testing material without resistance.The device does not articulate freely and smoothly.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have definitively contributed to the event.Therefore, no further clinical/medical assessment is warranted at this time.Should any additional clinically relevant medical information be provided, this case would be re-assessed.The complaint of difficult to open was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, an impact event inconsistent with normal use, wear from prolonged use, misuse or rough handling, or inadequate routine maintenance.The complaint of flaked was not confirmed and the root cause could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PCH BSKT DCKBL UPBTR 1.5MM
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15902601
MDR Text Key304684498
Report Number1219602-2022-01893
Device Sequence Number1
Product Code NBH
UDI-Device Identifier03596010023360
UDI-Public03596010023360
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number012014
Device Catalogue Number012014
Device Lot Number50992445
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/08/2022
Initial Date FDA Received12/02/2022
Supplement Dates Manufacturer Received01/26/2023
Supplement Dates FDA Received01/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/03/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-