Model Number 37612
Device Problems Failure to Interrogate (1332); Improper or Incorrect Procedure or Method (2017); Battery Problem (2885); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/30/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
It was reported that the patient electively turned the therapy off and stopped charging the ins in june of last year (for reasons unrelated to the device/therapy).About a month ago, the patient tried turning therapy back on, but their patient programmer (pp) indicated that it "could not find the battery," and the recharger was showing an information screen that indicated the ins was in a discharged state.The agent reviewed that the ins was likely overdischarged, and for that reason, the external devices were not communicating with it.The patient was redirected to their healthcare provider to address the issue further.The patient mentioned that this was the second time this has happened to their ins, noting that the previous time their hcp had to use the physician recharge mode so they could charge the ins again (the patient confirmed this was previously reported to pats).No symptoms/complications were reported.See (b)(4) (pe: (b)(4) for the previous overdischarge report.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the manufacturer¿s representative (rep) reported the healthcare provider (hcp) was going to see the patient on december 9th to try and perform a physician mode recharge.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (M
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15902999
MDR Text Key307893759
Report Number3004209178-2022-15821
Device Sequence Number1
Product Code MRU
UDI-Device Identifier00643169864238
UDI-Public00643169864238
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2019
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/30/2022
Initial Date FDA Received12/02/2022
Supplement Dates Manufacturer Received12/08/2022
Supplement Dates FDA Received12/13/2022
Date Device Manufactured09/13/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexMale
-
-