• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ANGIOJET SPIROFLEX; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ANGIOJET SPIROFLEX; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 10706
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2022
Event Type  malfunction  
Event Description
It was reported that catheter break occurred.An angiojet spiroflex was used for thrombectomy procedure.Upon preparation, the guide catheter was inserted and appeared to be cracked.The procedure was completed with another of the same device.There were no patient complications reported.
 
Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of the spiroflex thrombectomy.The pump, effluent/supply line, shaft, tip, piston and spike line were visually inspected.Visual examination that the shaft is kinked and the hypotube is separated 40.7cm from the tip.Microscopic examination revealed no additional damages.
 
Event Description
It was reported that catheter break occurred.An angiojet spiroflex was used for thrombectomy procedure.Upon preparation, the guide catheter was inserted and appeared to be cracked.The procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET SPIROFLEX
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15905072
MDR Text Key305885400
Report Number2124215-2022-49356
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/27/2022
Device Model Number10706
Device Catalogue Number10706
Device Lot Number0027027301
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/09/2022
Initial Date FDA Received12/02/2022
Supplement Dates Manufacturer Received12/08/2022
Supplement Dates FDA Received12/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-