• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. LEAD MODEL 302

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. LEAD MODEL 302 Back to Search Results
Model Number 302-20
Device Problems Low impedance (2285); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2022
Event Type  malfunction  
Event Description
It was reported that patient was seen in clinic for a programming adjustment.In clinic, the device showed low impedance.The patient did not feel any pain.X-rays have been completed.The patient recently had a generator replacement surgery a few months ago and it was noted that on the implant card the impedance was low, but still within normal limits at previous surgery.X-rays have not been reviewed by manufacturer to date.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
 
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
Physician later reported that there was no trauma or manipulation prior to low impedance.X-rays were reviewed by manufacturer.The x-rays were provided with report of low impedance.The generator placement was located below rib 4.Based on the images provided, the feedthrough wires were intact, and the pin appeared to be fully inserted.There is strain relief present and it is placed per labeling.Three tie downs are seen and therefore are not placed per labeling, as only 2 should be used.There was a portion of the lead that appeared to be routed behind the generator.There was a small section of the lead in the neck that appeared as anomaly with the possibility of a fracture, however due to the quality of the image being lost when zooming in, the fracture/discontinuity cannot be confirmed.The cause of the patient¿s low impedance could not be determined based on the images provided.Note that the presence of a micro-fracture and/or a lead discontinuity could not be ruled out in the portion of the lead that is not visible in the provided images.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
dana sprague
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key15905374
MDR Text Key307794927
Report Number1644487-2022-01544
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/13/2009
Device Model Number302-20
Device Lot Number27295
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 11/07/2022
Initial Date FDA Received12/02/2022
Supplement Dates Manufacturer Received12/08/2022
Supplement Dates FDA Received12/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/09/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexFemale
-
-