Model Number N/A |
Device Problem
Separation Failure (2547)
|
Patient Problem
Bone Fracture(s) (1870)
|
Event Date 11/07/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Device in the process of being returned.
|
|
Event Description
|
It was reported patient underwent a hip revision approximately 8 years post implantation due to a competitors non cemented stem loosening.During the procedure the arcos sts distal trial became stuck in femur.Surgeon had to perform an osteotomy to open the femur to remove the trial, he then had to cable the osteotomy after revision stem implanted.No additional information is available.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g3; h2; h3; h6 h6 component code: mechanical (g04) ¿ stem visual examination of the provided pictures identified the provisional has heavy gouging and damage.It is unknown if the damage was present prior to the removal by cutting.Lot identification is necessary for review of device history records; lot identification was not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional information on the reported event.
|
|