Catalog Number 383057 |
Device Problem
Degraded (1153)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/06/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the bd intima-ii¿ closed iv catheter system experienced a damaged catheter connector.The following information was provided by the initial reporter: the patient was hospitalized for intravenous administration, and a closed venous indwelling needle was used, and it was found that the connection was not smooth.
|
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd intima-ii¿ closed iv catheter system experienced a damaged catheter connector.The following information was provided by the initial reporter: the patient was hospitalized for intravenous administration, and a closed venous indwelling needle was used, and it was found that the connection was not smooth.
|
|
Manufacturer Narrative
|
H6: investigation summary: a device history review was conducted for lot number 1139438.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.
|
|
Search Alerts/Recalls
|