This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified, nor could the phenomenon be confirmed/reproduced.The event can be detected/prevented by following the instructions for use which state: [3.1 precautions for installation and connection].·properly and securely connect all cables.If the cable connector has connection screws, tighten up the screws.Otherwise, equipment damage or malfunction can result.·never apply excessive force to connectors.This could damage the connectors.Olympus will continue to monitor field performance for this device.
|