MEDTRONIC XOMED INC. TRICUT® STRAIGHT SHAFT 3.5MM 11CM LONG M4 ROTATABLE; BUR, EAR, NOSE AND THROAT
|
Back to Search Results |
|
Model Number 1883504HR |
Device Problem
Material Fragmentation (1261)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the blade tip broke soon after the operation started.There are no residuals inside the patient.Not patient impact.
|
|
Manufacturer Narrative
|
H3: product analysis found visually, the distal tip was broken 4.78 inches from the distal end of the inner hub when returned.There were biological contaminants on the tip, outside diameter of the outer tube, and proximal end of the inner hub.There were also striations on the proximal end of the inner shaft.Functional testing could not be performed due to the broken state of the device.In the returned condition, there was an out of specification condition that was related to the complaint (due to physical damage).H6: fdm b21, fdr c21 and fdc d16 codes no longer apply.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|