• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Ambient Noise Problem (2877)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/08/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.While testing the device, the abnormal noise was present about half-way through running the unit.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
The customer reported that his olympus high flow insufflation unit was producing an abnormal sound during a laparoscopic gastrectomy procedure.According to the initial reporter, the sound continued for a while, but eventually subsided, and the procedure was completed with the subject device.There was no harm to the patient as a result of this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, there were a lack of characters on the light-emitting diode (led) display.However, the definitive root cause of the led display issue could not be determined.Additionally, the abnormal noise was not confirmed.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD
3-1 okamiyama
odakura nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15914866
MDR Text Key308069932
Report Number3002808148-2022-04849
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/08/2022
Initial Date FDA Received12/05/2022
Supplement Dates Manufacturer Received12/23/2022
Supplement Dates FDA Received01/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-